Impedance

Pronunciation: im-PEED-ants

Electrical resistance measurement used to assess lead integrity and optimize stimulation parameters.

Full Definition

Impedance in neuromodulation refers to the opposition to electrical current flow through the electrode-tissue interface, measured in ohms. Lead impedance measurements are routinely performed to assess the integrity of the implanted system and optimize stimulation parameters. Normal impedance values vary depending on electrode design, surface area, and tissue characteristics, but typically range from 300-1500 ohms for spinal cord stimulation systems. Abnormally high impedance may indicate lead fracture or poor tissue contact, while very low impedance may suggest insulation breach or short circuit. Impedance monitoring is essential for device programming, troubleshooting, and long-term system maintenance.

Usage

Usage note: Note correct spelling with 'a' not 'e' in the middle; measured in ohms.

In Context

  • "Lead impedance values remained stable at 450 ohms throughout the six-month follow-up period." — Device interrogation report
  • "Elevated impedance readings suggested potential lead fracture requiring surgical exploration." — Troubleshooting assessment

Also known as

electrical resistance lead impedance

Contrasted with

conductance

Don't confuse with

impedence resistance voltage