magnet mode

A temporary device operating state activated by placing a magnet over an implanted cardiac device to disable certain functions.

Full Definition

Magnet mode is a special operating state triggered when a strong magnet is placed over an implanted cardiac device, causing temporary functional changes designed for emergency situations or specific procedures. In pacemakers, magnet application typically activates asynchronous pacing at a fixed rate, bypassing sensing functions and rate-responsive features. In ICDs, magnets usually disable anti-tachycardia therapies while maintaining bradycardia pacing support. The magnet response is programmable and can be disabled entirely in some devices. Magnet mode is commonly used during surgical procedures with electrocautery, emergency situations where device function must be predictable, or when device interference is suspected. The effect persists only while the magnet remains in place.

Usage

Usage note: Specify the duration of magnet application and resulting device behavior in procedural documentation.

In Context

  • "The pacemaker was placed in magnet mode during electrocautery to prevent inappropriate inhibition." — Surgical procedure note
  • "Magnet mode can be disabled in modern devices if the response is not clinically desired." — Device programming guide

Also known as

asynchronous mode magnet response

Contrasted with

demand mode

Don't confuse with

emergency mode safe mode